A multi-laboratory assessment of lupus anticoagulant assays performed on the ACL TOP 50 family for harmonized testing in a large laboratory network

Emmanuel J Favaloro, Soma Mohammed, Ronny Vong, Kent Chapman, Priscilla Swanepoel, Geoffrey Kershaw, Nancy Cai, Sarah Just, Lynne Connelly, Timothy Brighton, Leonardo Pasalic

Research output: Contribution to journalArticlepeer-review

7 Citations (Scopus)
42 Downloads (Pure)

Abstract

INTRODUCTION: Lupus anticoagulant (LA) testing is commonly performed within hemostasis laboratories, and the ACL TOP 50 family of instruments represent a new "single platform" of hemostasis instrumentation. Our aim was to evaluate these instruments and manufacturer reagents or alternatives for utility in LA testing.

METHODS: Comparative evaluations of LA testing using newly installed ACL TOPs 550 and 750 as well as comparative assessments with existing "reference," predominantly Stago, instrumentation, and reagents. Evaluations comprised both dilute Russell viper venom time (dRVVT) and activated partial thromboplastin time (APTT)-based assays. Establishment of normal reference ranges (NRR).

RESULTS: The HemosIL dRVVT-based assays showed good comparability with the existing Stago reference method (R > 0.9) and could be considered as verified as fit for purpose. A variety of APTT assays was additionally evaluated for LA utility, and we identified from the assessment good utility of a non-Werfen solution in Hyphen BioMed Cephen reagents. NRR were established based on ≥120 normal individual plasma samples.

CONCLUSION: This evaluation of LA reagents on ACL TOP 50 Family instruments identified overall acceptable performance of both dRVVT (Werfen solution) and APTT (non-Werfen solution) to enable harmonization of LA testing in our large network.

Original languageEnglish
Pages (from-to)654-665
Number of pages12
JournalInternational Journal of Laboratory Hematology
Volume44
Issue number3
Early online date01 Mar 2022
DOIs
Publication statusPublished - Jun 2022