Abstract

This paper reports on an evaluation of the introduction of a blood bank automation system (Ortho AutoVue® Innova) in a hospital blood bank by considering the performance and workflow as compared with manual methods. The turnaround time was found to be 45% faster than the manual method. The concordance rate was found to be 100% for both ABO/Rh(D) typing and antibody screening in both of the systems and there was no significant difference in detection sensitivity for clinically significant antibodies. The Ortho AutoVue® Innova automated blood banking system streamlined the routine pre-transfusion testing in hospital blood bank with high throughput, equivalent sensitivity and reliability as compared with conventional manual method.
Original languageEnglish
Pages (from-to)58-63
Number of pages6
JournalTransfusion and Apheresis Science
Volume53
Issue number1
DOIs
Publication statusPublished - 2015