Identification of factor inhibitors by diagnostic haemostasis laboratories: a large multi-centre evaluation

RCPA QAP in Haematology Haemostasis Committee

Research output: Contribution to journalArticlepeer-review

68 Citations (Scopus)

Abstract

We have assessed the proficiency of diagnostic haemostasis facilities to correctly identify coagulation factor abnormalities and inhibitors. Forty-two laboratories participating in the external Quality Assurance Program (QAP) conducted by the RCPA agreed to participate and were each sent a set of eight samples (each 3 x 1 ml) for evaluation. They were asked to blind test these samples for the presence or absence of inhibitors, and where identified, to perform further analysis (including specific inhibitor analysis). In order to make the exercise more challenging, in addition to true factor inhibitors, samples were provided that reflected potential pre-analytical variables that might arise and complicate inhibitor detection or lead to false inhibitor identification. In brief, the sample set comprised a true high level factor (F) V inhibitor, a true moderate level FVIII inhibitor (but sample was defibrinogenated), a true lupus anticoagulant (LA), a normal (but slightly aged) plasma sample, a normal serum sample, a normal EDTA sample, an oral anticoagulant/vitamin K deficiency sample, and a gross heparin ( approximately 10 U/ml) contaminated sample. Sixty-three percent of participants correctly identified the true FV inhibitor as such, although the reported range varied greatly [10 to >250 Bethesda units (BU/ml)] and 46% correctly identified the true FVIII inhibitor, despite the complication of the sample presentation, although the reported range also varied (7 to 64 BU/ml). Some laboratories either failed to identify the inhibitor present, or misidentified the inhibitor type. The LA, the oral anticoagulant/vitamin K deficiency, the normal serum sample, and the normal (aged) sample were also correctly identified by most laboratories, as was the absence of specific factor inhibitors in these samples. However, a small subset of laboratories incorrectly identified the presence of specific factor inhibitors in some of these samples. The heparin sample was also correctly identified by most (68%) laboratories. In contrast, the normal EDTA sample was misidentified as a FV and/or FVIII inhibitor by most (68%) laboratories, and only one laboratory correctly identified this as an EDTA sample. Thus, we conclude that although laboratories are able, in most cases, to identify the presence of true factor inhibitors, there is a large variation in identified inhibitor levels and there are also some significant errors in identification (i.e. false negatives and misidentifications). In addition, there is a significant false positive error rate where some laboratories will identify the presence of specific factor inhibitors where no such inhibitor exists (i.e. false positives).

Original languageEnglish
Pages (from-to)73-8
Number of pages6
JournalThrombosis and Haemostasis
Volume96
Issue number1
DOIs
Publication statusPublished - Jul 2006

Fingerprint

Dive into the research topics of 'Identification of factor inhibitors by diagnostic haemostasis laboratories: a large multi-centre evaluation'. Together they form a unique fingerprint.

Cite this