Personal Exposure Prescription Method Reduces Dose in Radiography

Xiaoming Zheng, Hsien Wen Chiang, Jung Hui Li, Hsien Jen Chiang, Li Han Lin

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Purpose To evaluate the effectiveness of an automatic, personalized exposure prescription method designed to reduce radiation dose during radiography examinations. Methods Using standard imaging parameters of average-sized patients, the authors measured individual body-part thicknesses or imaging regions of 116 patients (69 men, 47 women) and calculated each patient's exposure amount according to the thickness of the part or region. The data were used to develop each patient's personalized exposure prescription. Using the personalized exposure prescriptions, authors acquired chest images of the patients on a Carestream DRXRevolution mobile digital radiography system. Results All images acquired using the personalized exposure prescription method were satisfactory for diagnosis; exposure indexes were above 1300, a figure deemed acceptable for diagnosis by the manufacturer. The personalized exposure method reduced the amount of radiation each patient received. Discussion Variation of tube voltage alone can control patients' exposure levels; however, using the personalized exposure prescription method eliminates the need to use automatic exposure controls. Conclusion The personalized exposure prescription method is an effective tool for reducing radiation to patients during radiography as well as for eliminating dose creep.

Original languageEnglish
Pages (from-to)435-440
Number of pages6
JournalRadiologic Technology
Volume89
Issue number5
Publication statusPublished - 01 May 2018

Fingerprint Dive into the research topics of 'Personal Exposure Prescription Method Reduces Dose in Radiography'. Together they form a unique fingerprint.

  • Cite this

    Zheng, X., Chiang, H. W., Li, J. H., Chiang, H. J., & Lin, L. H. (2018). Personal Exposure Prescription Method Reduces Dose in Radiography. Radiologic Technology, 89(5), 435-440.